Cleaning and precautions for the probe of animal B-ultrasound machine
Cleaning of the probe of the animal B-ultrasound machine:
1. Disconnect the probe from the host, wipe off all coupling agents on the probe with a soft fabric, and then rinse with clean water.
2. Clean the probe with neutral soap and warm water. Wipe off any visible residue on the surface of the probe with a soft cotton ball, gauze, or other fabric. If the stain dries on the surface of the probe, a soft and damp brush (such as a toothbrush) can be used to clean it.
3. Use sufficient drinking water to clean the soap residue on the surface of the probe.
4. Air dry or wipe dry with soft fabric.
5. The depth of immersion of the probe into the liquid must not exceed the specified position to prevent liquid from seeping into the probe from the junction and damaging the probe.
Warning: Do not immerse the entire probe in water, otherwise it may cause damage to the probe and pose unexpected safety risks.
The probe of the animal B-ultrasound machine is placed at a depth of immersion in water
Disinfection of animal B-ultrasound probe:
After each use, the ultrasound probe can be disinfected with liquid chemical disinfectants. The degree of disinfection is directly related to the time of contact with disinfectants. Increasing the contact time has a better disinfection effect.
Precautions for using the probe of the animal B-ultrasound machine:
1) Before connecting or using the probe, it is necessary to ensure that the probe, probe cable, and connector are in normal condition (without damage, cracks, or detachment, etc.);
2) When loading and unloading probes, it is necessary to confirm that the host is in a shutdown state before proceeding with the operation, in order to avoid damaging the host and probe during live operation;
3) Be particularly careful when using the probe, do not collide with any hard objects, and ensure that the probe is not subjected to any impact, even a small impact may damage the probe;
4) The probe plug cannot be immersed in water.
5) After use, wipe the probe with a soft, dry cloth.
Attention: ultrasound probes are highly sensitive medical instruments and are easily damaged due to improper operation. Be careful when using and prevent damage when not in use. Do not immerse the probe into any liquid beyond the depth indicated in the probe instructions.
Never unplug or plug in the probe connector while the device is turned on to avoid damaging the probe and the host; Once the probe is connected to the host, do not unplug or plug it in randomly to avoid poor contact of the probe connector. The probe should avoid falling or colliding, and the manufacturer is not responsible for any risks caused by this.